SPECIFICATION:
Coldhead Type |
KDE412SA |
|
Lowest Temperature: |
< 3.5K |
|
Cooling Capacity (50Hz): |
First Stage |
Second Stage |
40W @ 45K---KDC6000V 35W@50K---KDC4000F |
1.25W @ 4.2K---DC6000V
0.8W@4.2K---KDC4000F |
|
Cooldown Time(2nd stage): |
< 60min.(4.2K) |
|
Weight: |
Coldhead |
Compressor |
18.5 kg |
118 kg |
|
Compressor Type: |
KDC6000V/KDC4000F |
|
Power Consumption(50Hz) |
Steady |
Cooldown |
6.5kW |
7.2 kW |
|
Cooling Type: |
Water |
|
Cooling Water Requirement: |
﹥ 7 L/min |
|
Standard Flexline: |
20A×20m |
|
Maintenance Interval: |
Coldhead |
|
18 months |
Environmental Requirement
Item |
Operating |
Storage |
Ambient Temperature |
4-40 ℃ |
-20-65 ℃ |
Relative Humidity |
30%-70% |
10%-90% (Requiring No-condensing) |
Ambient atmospheric pressure |
70kPa~110kPa |
20kPa~110kPa |
Outline Drawing:
Typical Load Map: