Cryogenic refrigerator

Open
Home > Product > Cryogenic refrigerator

KDE412SA(4K)

2018-05-18
    KDE412SA Cryocooler consists of a Cylinder, No.1 Displacer, No.2 Displacer, driver mechanism,and Drive Motor. No.1 Displacer is connected to the Scotch Yoke which can be driven by the Drive Motor through the Crank with Bush so that the rotation of the Drive Motor can be varied to reciprocating motion of Scotch Yoke and Displacers. 

Specification:

Coldhead Type KDE412SA
Lowest Temperature  3.5K
Cooling Capacity (50Hz) First Stage Second Stage
40W @ 45K 1.25W @ 4.2K
Cooldown Time2nd stage):  60min.(4.2K)
Weight Coldhead Compressor
18.5 kg 118 kg
Compressor Type KDC6000V
Power Consumption(50Hz) Steady Cooldown
6.5kW 7.2 kW
Cooling Type: Water
Cooling Water Requirement  7 L/min
Standard Flexline 20A×20m
Maintenance Interval Coldhead
18 months

Environmental Requirement:

Item Operating Storage
Ambient Temperature 4-40  -20-65 
Relative Humidity 30%-70% 10%-90%
(Requiring No-condensing)
Ambient atmospheric pressure 70kPa~110kPa 20kPa~110kPa

CSIC Pride(Nanjing)Cryogenic Technology Co., Ltd.